lichtzen-ae logo.gif

 

 

 

 

     LICHTZEN AE     AITNIX 

 

 

 

 

 

 

 

  Thinking high tech  

 

 

About us  |  E-mail  |  Board  |  Site Map

 

   

 

 

  Home > Instruments > 광 계측기기
img53.gif
  광 계측기기 (OPTICAL INSTRUMENTS)  

 

 

 

 

 

 

 
 

IL1700 RESEARCH RADIOMETER

 

 

 

 

 

  • 10 Decade Dynamic Range
  • 0.1 Percent Linearity
  • 40 Decade Integration Range
  • Light Integration up to 18 Years
  • 10 Calibration Factors in Memory
  • Direct Reading in any Units
  • Remote Control of Functions
  • Darkroom Hold Reading Feature
  • Selectable Reverse Bias (0 or 5 V)
  • 115/230 VAC, 8-15 VDC, Battery Power
  • Built-In NiCad Battery Recharger
  • Analog Recorder Output (0-1 volt)
  • RS232C and TTL Digital Outputs
  • Automatic Zeroing
  • Automatic 100 Percent
  • Automatic Ranging (On/Off)

 

FRONT PANEL CONTROLS

REAR PANEL

The 'ZERO' button nulls the ambient signal to zero. The 'D.C.' button selects direct measurement mode. The 'INT' button selects light dose integration mode. The 'HOLD' button freezes the display and sends data to the optional printer.

The 'AUTO RANGE' button selects either automatic or fixed range. The 'AUTO 100%' button sets the displayed signal to 100 percent.

The '5 V BIAS' button provides an additional 5 volt reverse bias for flash integrations and for all vacuum photodiode detectors.

The 'DATA/FACTOR DISPLAY' button selects either detector output or sensitivity factor display.

The 'FACTOR SELECT' button increments through the 10 sensitivity factor memory registers.

The 'MSD', 'LSD' and 'EXP' buttons allow the user to adjust the sensitivity factor.

The 'POWER' button turns the power on/off.

An AUTO SHUT-OFF feature automatically turns off the instrument if idle for more than 10 minutes in battery mode.

'BAT' and 'AC' LED's indicate the power source, with low battery warning.

The 'CHG' LED indicates batteries recharging.

The 'OUTPUT RECORDER/TTL' port allows simultaneous autoranging analog recorder output (0-1 volt) and TTL digital data (0-5 volts).

The 'AC VOLTAGE' switch selects 115 or 230 VAC.

The 'COMPUTER/PRINTER' switch configures the standard 'RS232C DIGITAL OUTPUT' connector to match DTE or DCE equipment.

The 'SOURCE AC/BAT' switch selects AC or battery power.

The '90-130/180-240 VAC' power input plug accepts the provided AC power cable.

The detector cable plugs into the 'SENSOR INPUT', which also provides voltages of +15 V, 5 V, and -9 V for various sensor probes.

The 24 line, card edge connector labeled 'ACCESSORY INPUT' allows various input/output functions: auxiliary power input, recorder output, RAM voltage check for constant memory, and remote operation of 'ZERO', 'D.C.', 'INT', and 'HOLD' functions.

 

 

ELECTRICAL / MECHANICAL:

Dynamic Range: 2 x 10-13 to 2 x 10-3 A.
Min. Resolution: 0.01 x 10-11 Amperes.
Linearity: ±0.1%, top 8 decades; ±0.2%, ±1 digit, bottom 2 decades.
Display: 3 1/2 digit + 1 1/2 digit exp.
Display Range: 1.000 x 10-19 to 9.99 x 10+19 display units.
Input Impedance: Virtual Ground (op-amp summing node).
AC Power: 90-130 / 180-240 V, 50-400 Hz.
DC Power: 8-15 V external, or 6 internal 'C' cells.
Max Power: 11 Watts.
RS232C Output: DTE or DCE, ± 8 V.
TTL Output: 0 to 5 V.
Recorder Output: 408 Ω impedance, 100 to 1000 mV, ranging every decade, current limited.
Size: 90 mm x 220 mm x 240 mm (3.5" x 8.7" x 9.4").
Weight: 2.3 kg (5.1 lbs.).

 

 

 

광량계, IL1700, IL1400, uv curing, radiometer, 경화, 자외선, 적외선, 근적외선, 원적외선, 가시광선, 건조, 화학방사선, 세정, 살균, 수처리, 태양광, 복사량측정, 광량계, 조도계, 광도계, 센서, 스펙트로메타, 라디오메타, 포토메타, 스펙트로라디오메타, uv, LED, 자외선측정, UV광량, IL393, IL290, LZ393, IL393, ILT400, ILT490, ILT1700, ILT1400, ILT900, ILT900 LED, 색차계,조사, 분광, 스펙트럼, 분석, 복사도, 환경, 생화학, 광생물학, 레이저,

ultraviolet radiation, uv meter, light bugs, UVA, UV-b, UV-A, UVB, UVC, UV-Cil290a, il390c, il393, uv exposure,

노광, spot irradiance measurements, il1747, il1447, il1745, il1445, pir irradiance calibration,

photometry, photometer system, photometric illuminance, flash photometer, radiometer, high gain lens, flash sources, flash effective intensity, il1710, il1410, il1711, il1715, radiometer, visible, near-ir, irradiance measurements, radiometry, spatial response, uv detection, radiance measurements, irradiance detector, nist traceable, il1705, il1405, il1715, photometer system, photoresist, a response, b response, uv measurement, blue visible irradiance, exposing photoresist, uva, photoresist equipment, radiometer, uv sensitivity, il1740a

 

Top of Page

 

 

   

DTX logo

경기도 시흥시 정왕동 1288-2 번지 동우디지털파크 A-512 |   E-mail : contact@dtx.co.kr  

Tel : 031-497-1134 (代)   |  Fax : 031-497-1138   |  All Rights Reserved, Copyright (C) DTX, INC. 2008

Tel : 031-419-6770 (영업부 직통)   |   Internet Fax : 0505-497-7710