  환경 분석 기기 (ENVIRONMENTAL INSTRUMENTS)

 

 

 

 

 

가스분석기 컨버터 유닛 (GAS ANALYZER PRIMARY CONVERTER UNIT)   

 

 

 

 MGL - 19M-1(CO) electrochemical stationary

 

 기술사양 (GAS ANALYZER MGL-19M.1 PRIMARY CONVERTER UNIT)

CO measurable concentration range

: 0 - 200 mg/m3

Zero drift, less than

: 0.5 ppm/30days

Calibration drift, less than

: 5%/30days

Linearity error

: 1% of full scale

BPP Power supply voltage

: ~ 220V, 50 Hz

Instrument recourse, not less

: 10000 hours

BPP Dimensions

: 85 x 100 x 105 mm

BPP Weight

: 0,4 kg

* The resource is determined without the consideration of the electrochemical sensor change.. 

 

 

 운용조건 (OPERATION CONDITIONS)

Ambient temperature

: +5 - +40° С

Pressure

: 630 to 800 mmHg

Humidity

: Up to 98%

Sample

: diffusion

 

 기능 (ASSIGNMENT)

The unit of primary converter (BPP) of MGL-19M-1 is assigned for carbon oxide (CO) continuous measurements in non hazardous area in the working zone air.
BPP of MGL-19M-1 satisfies all the requirements of Gosgortehnadsor 12-341-00.
The output linear current signal is established in the range 4 - 20 or 0 - 5 mA
(according to customer’s preliminary choice).
Maximal distance from BPP to secondary indicating devices (data process unit or ammeter) does not exceed 500 m.
The analyzable component concentration is defined by the calculation way when the BPP is operating (mg/m3 or % vol.)

 

 운용원리 (THE OPERATION PRINCIPLE)

The electrochemical method of detection is in the basis of analyzer operation. Electrochemical sensor is used as a sensitive element. Measurable gas enters the sensor by diffusion and causes electrical current in electrodes, proportional to the gas concentration. This current proceeds through loaded resistor. Voltage from load resistor is amplified, standardized and comes in voltage-current transformer. The current value is an analytical signal for measurement of analyzed gas concentration.

 

 DEVICE FUNCTIONALITIES AND FEATURES

The instrument is executed in plastic case with a transparent cover (dust, water and blow resist perfomance). It has a fixing elements for wall mounting, light emitting diode (LED), showing the power feed presence (~ 220 V, 50 Hz) and two LEDs to indicate the switching-on of the alarm system analyzable component established levels. The keys "Калибровка", "+", "-" и "Уст.0" are necessary to carry out the analyzer calibrating. There is the key of the sound alarm switching-off on the lateral panel (“Сброс”).

BPP of "MGL-19M" model has two levels of alarm system switching-on, established on the factory: “порог 1” (level 1) corresponds the analyzable gas extreme allowable concentration (ПДК) in the working zone air СO -20 mg/m3, “порог 2” corresponds the 5 values of analyzable gas ПДК:

Alarm relay contacts provides the commuted devices connection (sound and light alarm, the control of external executive device by means of the relay contacts). Maximal commuted current is 2 A, voltage – 120 V.

Calibration interval is 6 months
Main calibration interval – 1 year.

 

포터블가스분석기, 황화수소분석기, 잔류가스분석기,오존분석기,O3, 염소분석기, 암모니아분석기, 벤젠분석기, 나프탈렌분석기,배출가스분석기, 대기가스분석기 방향족 분석기, 염기산화물분석기, 금속이온분석기, 바나듐분석기, 구리이온분석기, 수소분석기, 황산화물분석기, 색상분석기, 톨루엔분석기, 코발트분석기,CIP분석기, MEG분석기, 페놀분석기, 광화이버 스펙트로메타, 샘플링시스템, 벤젠, 자일렌, 발전소, 소각로, NOX, SOX, Process analyzers, Process stream,OZONE ANALYZER, Gas, liquid,  Hydrogen sulfide analyzer, Tail gas analyzer for the Claus sulfur recovery process, UV / VIS / NIR Diode array spectrophotometer, Chlorine analyzer, Ammonia analyzer, Aromatics analyzer, chlorine dioxide analyzer, Metal ions analyzer, MEG analyzer,  TiCl4 ,vanadium analyzer, CIP analyzer, Copper ions analyzer, Hydrogen analyzer, Sulfur dioxide analyzer, TBC analyzer, Color analyzer, Fiber optic spectrometers, Sampling systems, Cl2, BTX, Benzene, Toluene, Xylene, MEHQ, NH3, Multi Components, Concentration monitoring, On-line, Continuous measurements, Cobalt recovery, Sulfur recovery, Air demand, Ammonia slip, power plant, SO2, COS, CS2, Sulfur, H2S, NCl3, ClO2, O3"

 