기술사양 (TECHNICAL SPECIFICATION) : OZONE ANALAZER 3-02.P-R
|
|
|
The range of ozone measuring concentrations |
:
0 - 500 mcg/m3 |
|
|
Zero drift, less than |
:
1 ppb / 30 days |
|
|
Calibration drift, less than |
:
5% / 30 days |
|
|
Linearity error |
: 2 % of full scale |
|
|
Output signal: analogue (additionally) digital |
:
0 - 5 mА or : 4 - 20 mА :
RS-232 |
|
|
Dimensions |
:
145 х 270 х 390 mm |
|
|
Weight, not more |
:
5,5 kg |
|
|
The limit of final value stabilization period То,9д , not
more |
:
30 sec |
|
|
Power supply |
:
220V, 50 Hz |
|
|
Power consumption |
: 15 VA
|
|
|
|
운용조건 (THE AMBIENT OPERATION CONDITIONS)
|
|
|
Temperature |
:
+10 ...+35оС |
|
|
Pressure |
:
630 - 800 mm Hg. |
|
|
Humidity |
:
Up to 95% |
|
|
Parameters of the analyzed ambient air in the gas analyzer input
: |
:
- |
|
|
Temperature |
:
+10...+40оС |
|
|
Volume flow |
:
1,5 ± 0,2 l/min |
|
|
relative humidity |
:
Up to 98% |
|
기능 (ASSIGNMENT)
Because of low dimensions and high reliability this instrument can be used
for ozone background monitoring in inhabited zone, for ozone monitoring in
different process working zone when concentrations are near extreme marginal
values. This analyzer is a basis model of chemiluminescent ozone meters,
manufactured in the enterprise. Analyzer is a main resort of measuring in the
level of extreme marginal values in Russia in present.
적용분야 (APPLICATIONS)
Industry ozone monitoring in working zone air.
운용원리 (PRINCIPLE OF OPERATION)
Gas analyzer is a gas comparator providing high linearity of signal
transformation function. The comparison is carried out between a model mixed gas
from calibrator and analyzable gas. The basis of analyzer 3-02 П1 operation is
an effect of the heterogeneous chemiluminescence, which is carried out due to
ozone exothermic reaction with composition oxidizable chemical substances. The
intensity of a composition luminescence is proportional to density of ozone in a
mixed gas. Intensity is measured and transformed to the digital signal,
indicated in the analyzer display. Built-in micropump provides the feed of
analyzable gas sample in chemiluminescent reactor.
플로우 차트 (GAS ANALYZER FLOWCHART)

|